NI FGEN Frequency List Output Functions

Palash Khare edited this page Jun 8, 2021 · 3 revisions

Frequency List Output Functions

niFgen_CreateFreqList

ViStatus niFgen_CreateFreqList (ViSession vi, ViInt32 waveform, ViInt32 frequencyListLength, ViReal64[] frequencyArray, ViReal64[] durationArray, ViInt32* frequencyListHandle);

Purpose

Creates a frequency list from an array of frequencies (frequencyArray) and an array of durations (durationArray). The two arrays should have the same number of elements, and this value must also be the size of the frequencyListLength. The function returns a handle that identifies the frequency list (the frequencyListHandle). You can pass this handle to niFgen_ConfigureFreqList to specify what frequency list you want the signal generator to produce.

Note:  The signal generator must not be in the Generating state when you call this function.

You must call the niFgen_ConfigureOutputMode function to set the outputMode parameter to NIFGEN_VAL_OUTPUT_FREQ_LIST before calling this function. A frequency list consists of a list of frequencies and durations. The signal generator generates each frequency for the given amount of time and then proceeds to the next frequency. When the end of the list is reached, the signal generator starts over at the beginning of the list.

Parameters

Input
Name Type Description
vi ViSession Identifies your instrument session. vi is obtained from the niFgen_init, niFgen_InitWithOptions, or niFgen_InitializeWithChannels functions and identifies a particular instrument session.
waveform ViInt32

Specifies the standard waveform that you want the signal generator to produce. NI-FGEN sets the NIFGEN_ATTRIBUTE_FUNC_WAVEFORM attribute to this value.

Default Value: NIFGEN_VAL_WFM_SINE

Defined Values

NIFGEN_VAL_WFM_SINE Specifies that the signal generator produces a sinusoid waveform.
NIFGEN_VAL_WFM_SQUARE Specifies that the signal generator produces a square waveform.
NIFGEN_VAL_WFM_TRIANGLE Specifies that the signal generator produces a triangle waveform.
NIFGEN_VAL_WFM_RAMP_UP Specifies that the signal generator produces a positive ramp waveform.
NIFGEN_VAL_WFM_RAMP_DOWN Specifies that the signal generator produces a negative ramp waveform.
NIFGEN_VAL_WFM_DC Specifies that the signal generator produces a constant voltage.
NIFGEN_VAL_WFM_NOISE Specifies that the signal generator produces white noise.
NIFGEN_VAL_WFM_USER Specifies that the signal generator produces a user-defined waveform as defined with the niFgen_DefineUserStandardWaveform function.
Name Type Description
frequencyListLength ViInt32

Specifies the number of steps in the frequency list you want to create. The value must be between the minimum and maximum frequency list lengths that the signal generator allows. You can obtain the minimum and maximum frequency list lengths from the minimumFrequencyListLength and maximumFrequencyListLength parameters in the niFgen_QueryFreqListCapabilities function.

frequency and duration must each be at least as long as this frequency list length.

Default Value: None

frequencyArray ViReal64[]

Specifies the array of frequencies to form the frequency list. The array must have at least as many elements as the value you specify in frequencyListLength. Each frequencyArray element has a corresponding durationArray element that indicates how long that frequency is repeated.

Units: hertz

Default Value: None

durationArray ViReal64[]

Specifies the array of durations to form the frequency list. The array must have at least as many elements as the value that you specify in frequencyListLength. Each durationArray element has a corresponding frequencyArray element and indicates how long in seconds to generate the corresponding frequency.

Units: seconds

Default Value: None

Output
Name Type Description
frequencyListHandle ViInt32* Returns the handle that identifies the new frequency list. You can pass this handle to niFgen_ConfigureFreqList to generate the arbitrary sequence.

Return Value

Name Type Description
status ViStatus Returns the status code of this operation. The status code either indicates success or describes an error or warning condition. You can examine the status code from each call to an NI-FGEN function to determine if an error occurred. To obtain a text description of the status code, call the niFgen_error_message function. To obtain additional information about the error condition, call the niFgen_GetError function. To clear the error information from NI-FGEN, call the niFgen_ClearError function.

The general meaning of the status code is as follows:
Value Meaning
0 Success
Positive Values Warnings
Negative Values Errors

niFgen_ConfigureFreqList

ViStatus niFgen_ConfigureFreqList (ViSession vi, ViConstString channelName, ViInt32 frequencyListHandle, ViReal64 amplitude, ViReal64 dcOffset, ViReal64 startPhase);

Purpose

Configures the attributes of the signal generator that affect frequency list generation (the NIFGEN_ATTRIBUTE_FREQ_LIST_HANDLE, NIFGEN_ATTRIBUTE_FUNC_AMPLITUDE, NIFGEN_ATTRIBUTE_FUNC_DC_OFFSET, and NIFGEN_ATTRIBUTE_FUNC_START_PHASE attributes).

Note:  The signal generator must not be in the Generating state when you call this function.

You must call the niFgen_ConfigureOutputMode function to set the outputMode parameter to NIFGEN_VAL_OUTPUT_FREQ_LIST before calling this function.

Parameters

Input
Name Type Description
vi ViSession Identifies your instrument session. vi is obtained from the niFgen_init, niFgen_InitWithOptions, or niFgen_InitializeWithChannels functions and identifies a particular instrument session.
channelName ViConstString

Specifies the channel name for which you want to configure the frequency list.

Default Value: "0"

frequencyListHandle ViInt32 Specifies the handle of the frequency list that you want the signal generator to produce. NI-FGEN sets the NIFGEN_ATTRIBUTE_FREQ_LIST_HANDLE attribute to this value. You can create a frequency list using the niFgen_CreateFreqList function, which returns a handle that you use to identify the list.

Default Value: None
amplitude ViReal64

Specifies the amplitude of the standard waveform that you want the signal generator to produce. This value is the amplitude at the output terminal. NI-FGEN sets the NIFGEN_ATTRIBUTE_FUNC_AMPLITUDE attribute to this value.

For example, to produce a waveform ranging from –5.00 V to +5.00 V, set the amplitude to 10.00 V.

Units: peak-to-peak voltage

Default Value: None

Note:   This parameter does not affect signal generator behavior when you set the waveform parameter of the niFgen_ConfigureStandardWaveform function to NIFGEN_VAL_WFM_DC.

dcOffset ViReal64

Specifies the DC offset of the standard waveform that you want the signal generator to produce. The value is the offset from ground to the center of the waveform you specify with the waveform parameter, observed at the output terminal. For example, to configure a waveform with an amplitude of 10.00 V to range from 0.00 V to +10.00 V, set the dcOffset to 5.00 V. NI-FGEN sets the NIFGEN_ATTRIBUTE_FUNC_DC_OFFSET attribute to this value.

Units: volts

Default Value: None

startPhase ViReal64 Specifies the horizontal offset of the standard waveform you want the signal generator to produce. Specify this attribute in degrees of one waveform cycle. NI-FGEN sets theNIFGEN_ATTRIBUTE_FUNC_START_PHASE attribute to this value. A start phase of 180 degrees means output generation begins halfway through the waveform. A start phase of 360 degrees offsets the output by an entire waveform cycle, which is identical to a start phase of 0 degrees.

Units: degrees of one cycle

Default Value: None degrees

Note:   This parameter does not affect signal generator behavior when you set the waveform parameter to NIFGEN_VAL_WFM_DC.

Return Value

Name Type Description
status ViStatus Returns the status code of this operation. The status code either indicates success or describes an error or warning condition. You can examine the status code from each call to an NI-FGEN function to determine if an error occurred. To obtain a text description of the status code, call the niFgen_error_message function. To obtain additional information about the error condition, call the niFgen_GetError function. To clear the error information from NI-FGEN, call the niFgen_ClearError function.

The general meaning of the status code is as follows:
Value Meaning
0 Success
Positive Values Warnings
Negative Values Errors

niFgen_ClearFreqList

ViStatus niFgen_ClearFreqList (ViSession vi, ViInt32 frequencyListHandle);

Purpose

Removes a previously created frequency list from the signal generator memory and invalidates the frequency list handle.

Note:   The signal generator must not be in the Generating state when you call this function.

Parameters

Input
Name Type Description
vi ViSession Identifies your instrument session. vi is obtained from the niFgen_init, niFgen_InitWithOptions, or niFgen_InitializeWithChannels functions and identifies a particular instrument session.
frequencyListHandle ViInt32

Specifies the handle of the frequency list you want the signal generator to remove. You create multiple frequency lists using niFgen_CreateFreqList. niFgen_CreateFreqList returns a handle that you use to identify each list. Specify a value of -1 to clear all frequency lists.

Defined Value

NIFGEN_VAL_ALL_FLISTS — Remove all frequency lists from the signal generator.

Default Value: None

Return Value

Name Type Description
status ViStatus Returns the status code of this operation. The status code either indicates success or describes an error or warning condition. You can examine the status code from each call to an NI-FGEN function to determine if an error occurred. To obtain a text description of the status code, call the niFgen_error_message function. To obtain additional information about the error condition, call the niFgen_GetError function. To clear the error information from NI-FGEN, call the niFgen_ClearError function.

The general meaning of the status code is as follows:
Value Meaning
0 Success
Positive Values Warnings
Negative Values Errors

niFgen_QueryFreqListCapabilities

ViStatus niFgen_QueryFreqListCapabilities (ViSession vi, ViInt32* maximumNumberOfFreqLists, ViInt32* minimumFrequencyListLength, ViInt32* maximumFrequencyListLength, ViReal64* minimumFrequencyListDuration, ViReal64* maximumFrequencyListDuration, ViReal64* frequencyListDurationQuantum);

Purpose

Returns the attributes of the signal generator that are related to creating frequency lists. These attributes are NIFGEN_ATTRIBUTE_MAX_NUM_FREQ_LISTS, NIFGEN_ATTRIBUTE_MIN_FREQ_LIST_LENGTH, NIFGEN_ATTRIBUTE_MAX_FREQ_LIST_LENGTH, NIFGEN_ATTRIBUTE_MIN_FREQ_LIST_DURATION, NIFGEN_ATTRIBUTE_MAX_FREQ_LIST_DURATION, and NIFGEN_ATTRIBUTE_FREQ_LIST_DURATION_QUANTUM.

Parameters

Input
Name Type Description
vi ViSession Identifies your instrument session. vi is obtained from the niFgen_init, niFgen_InitWithOptions, or niFgen_InitializeWithChannels functions and identifies a particular instrument session.
Output
Name Type Description
maximumNumberOfFreqLists ViInt32 Returns the maximum number of frequency lists that the signal generator allows. NI-FGEN obtains this value from the NIFGEN_ATTRIBUTE_MAX_NUM_FREQ_LISTS attribute.
minimumFrequencyListLength ViInt32 Returns the minimum number of steps that the signal generator allows in a frequency list. NI-FGEN obtains this value from the NIFGEN_ATTRIBUTE_MIN_FREQ_LIST_LENGTH attribute.
maximumFrequencyListLength ViInt32 Returns the maximum number of steps that the signal generator allows in a frequency list. NI-FGEN obtains this value from the NIFGEN_ATTRIBUTE_MAX_FREQ_LIST_LENGTH attribute.
minimumFrequencyListDuration ViReal64 Returns the minimum duration that the signal generator allows in a step of a frequency list. NI-FGEN obtains this value from the[NIFGEN_ATTRIBUTE_MIN_FREQ_LIST_DURATION attribute.
maximumFrequencyListDuration ViReal64 Returns the maximum duration that the signal generator allows in a step of a frequency list. NI-FGEN obtains this value from the NIFGEN_ATTRIBUTE_MAX_FREQ_LIST_DURATION attribute.
frequencyListDurationQuantum ViReal64 Returns the quantum of which all durations must be a multiple in a frequency list. NI-FGEN obtains this value from the NIFGEN_ATTRIBUTE_FREQ_LIST_DURATION_QUANTUM attribute.

Return Value

Name Type Description
status ViStatus Returns the status code of this operation. The status code either indicates success or describes an error or warning condition. You can examine the status code from each call to an NI-FGEN function to determine if an error occurred. To obtain a text description of the status code, call the niFgen_error_message function. To obtain additional information about the error condition, call the niFgen_GetError function. To clear the error information from NI-FGEN, call the niFgen_ClearError function.

The general meaning of the status code is as follows:
Value Meaning
0 Success
Positive Values Warnings
Negative Values Errors

Table of Contents

Internal Development

Creating and Setting Up a gRPC Server

Server Security Support

Creating a gRPC Client

gRPC Client Examples

Session Utilities API Reference

Driver Documentation

gRPC API Differences From C API

Sharing Driver Sessions Between Clients

C API Docs
NI-DAQmx
NI-DCPOWER
NI-DIGITAL PATTERN DRIVER
NI-DMM
NI-FGEN
NI-FPGA
NI-RFmx Bluetooth
NI-RFmx NR
NI-RFmx WCDMA
NI-RFmx GSM
NI-RFmx CDMA2k
NI-RFmx Instr
NI-RFmx LTE
NI-RFmx SpecAn
NI-RFmx TD-SCDMA
NI-RFmx WLAN
NI-RFSA
NI-RFSG
NI-SCOPE
NI-SWITCH
NI-TCLK
NI-XNET
Clone this wiki locally